Fluke 106 Digital Multimeter-i400E Clamp Bundle
Fluke 106 Digital Multimeter-i400E Clamp Bundle Digital Multimeter  Fluke Selangor, Penang, Malaysia, Kuala Lumpur (KL), Petaling Jaya (PJ), Butterworth Supplier, Suppliers, Supply, Supplies | MOBICON-REMOTE ELECTRONIC SDN BHD
Quantity
The Fluke 106 digital multimeter is made to fit the way you work! It is designed to fit in the palm of your hand and go with you no matter where your job takes you.
 

The Fluke i400E AC Current Clamp is an ideal tool to extend current measurement. It is designed to work with the new Fluke 101, 106 and 107 palm sized digital multimeters, as well as most Fluke multimeters or other voltage measurement tools that accepts 19.05 mm shrouded banana plugs and has mV ac measurement functionality
 

Fluke 106 Palm-sized Digital Multimeter

  • Voltage, resistance, continuity, capacitance
  • Input terminal for ac and dc current measurements to 10 A and current
  • Data hold

Fluke i400E AC Current Clamp

  • Compatible with most Fluke multimeters or mV ac measurement tools with these features:
    • 400 mV ac range
    • 19.05 mm banana jack inputs
    • Input accuracy of 2 % or greater to take full advantage of the accuracy of the product
    • Input impedance of >1 MΩ
  • 1 mV/A ac output guarantees easy current readings on you meter
  • CAT III 600 V safety rating
  • Specially designed to offer maximum utility in a very compact shape
  • Take accurate current readings

Fluke 106 Palm-sized Digital Multimeter

Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C, relative humidity at 0 % to 75 %. Accuracy specifications take the form of: ±([% of Reading] + [Number of Least Significant Digits])


Accuracy Specifications
Function Range Resolution Accuracy
AC Volts1 6.000 V
60.00 V
600.0 V
0.001 V
0.01 V
0.1 V
1.0 % + 3
DC Volts 6.000 V
60.00 V
600.0 V
0.001 V
0.01 V
0.1 V
0.5 % + 3
AC Millivolts 600.0 mV 0.1 mV 3.0 % + 3
Diode Test2 2.000 V 0.001 V 10 %
Resistance 
(Ohms)
400.0 O
4.000 kO
40.00 kO
400.0 kO
4.000 MO
40.00 MO
0.1 O
0.001 kO
0.01 kO
0.1 kO
0.001 MO
0.01 MO
0.5 % + 3
0.5 % + 2
0.5 % + 2
0.5 % + 2
0.5 % + 2
1.5% + 3
Capacitance3 50.00 nF
500.0 nF
5.000 µF
50.00 µF
500.0 µF
1000 µF
0.01 nF
0.1 nF
0.001 µF
0.01 µF
0.1 µF
1 µF
2 % + 5
2 % + 5
5 % + 5
5 % + 5
5 % + 5
5 % + 5
Frequency4 Hz
(10 Hz – 100 kHz)
50.00 Hz
500.0 Hz
5.000 kHz
50.00 kHz
100.0 kHz
0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz
0.1 kHz
NA
Duty Cycle4 1 % to 99 % 0.1 % NA
AC Current
(40 Hz to 200 Hz)
4.000 A
10.00 A
0.001 A
0.01 A
1.5 % + 3
DC Current 4.000 A
10.00 A
0.001 A
0.01 A
1.5 % + 3
  1. All AC, Hz, and duty cycle are specified from 1 % to 100 % of range. Inputs below 1 % of range are not specified.
  2. Typically, open circuit test voltage is 2.0 V and short circuit current is <0.6 mA.
  3. Specifications do not include errors due to test lead capacitance and capacitance floor (may be up to 1.5 nF in the 50 nF range).
  4. All AC, Hz, and duty cycle are specified from 1 % to 100 % of range. Inputs below 1 % of range are not specified.
Function Overload Protection Input Impedance (Nominal) Common Mode Rejection Ratio Normal Mode Rejection Ratio
AC Volts 600 V 1 >10 MO <100 pF 2 >60 dB at dc,
50 Hz or 60 Hz
-
AC Millivolts 600 mV >1M, <100 pF >80 dB at 50 Hz or 60 Hz -
DC Volts 600 V 1 >10 MO <100 pF >100 dB at dc,
50 Hz or 60 Hz
>60 dB at 50 Hz or 60 Hz
  1. 6 x 105 V Hz Max
  2. For mV (AC), input impedance is approximately 1 MO.
General Specifications
Maximum Voltage Between any Terminal and Earth Ground 600 V
Display (LCD) 6000 counts, updates 3/sec
Battery Type 2 AAA, NEDA 24A, IEC LR03
Battery Life 200 hours minimum
Temperature
Operating 0 °C to 40 °C
Storage -30 °C to 60 °C
Relative Humidity
Operating Humidity Non-condensing when <10°C
=90 % at 10 °C to 30 °C; =75 % at 30 °C to 40 °C
Operating Humidity, 40 MO Range =80 % at 10 °C to 30 °C; =70 % at 30 °C to 40 °C
Altitude
Operating 2000 m
Storage 12,000 m
Temperature Coefficient 0.1 X (specified accuracy) /°C (<18 °C or >28 °C)
Fuse Protection for Current Inputs 11A, 1000V Fast Fuse, Fluke specified part only
 
Size (HxWxL) 142 mm x 69 mm x 28mm
Weight 200 g
IP Rating IEC 60529: IP 40
Safety IEC 61010-1: 600 V CAT III, Pollution Degree 2
Electromagnetic Environment IEC 61326-1: Portable
Electromagnetic Compatibility Applies to use the Korea only.  Class A Equipment (Industrial Broadcasting and Communication Equipment)1
  1. This product meets requirements for industrial (Class A) electromagnetic wave equipment and seller or user should take notice of it. This equipment is intended for use in business environments and is not to be used in homes.

Fluke i400E AC Current Clamp

Electrical specifications
AC current range

1 A to 400 A

Output signal 1 mV/A ac
Accuracy 2.0 % ± 5 digits (45 Hz to 65 Hz)
2.5 % ± 5 digits (65 Hz to 400 Hz)
Typical bandwith 5 Hz to 20 kHz
Working voltage 600 V ac rms, in compliance with IEC 61010-1 and IEC 61010-2-030
Load impedance 1 A to 400 A
Maximum non-destructive current 600 A
Jaw inner diameter 30 mm
Maximum wire size 600 MCM
Standard plugs 19.04 mm dual banana plugs
Safety specifications
Category rating IEC 61010-1 and IEC 61010-2-030/IEC 61010-2-031 and
IEC 61010-2-032: 600 V CAT III
Pollution Degree 2
Electromagnetic environment IEC 61326-1: Portable
General Specifications
Output cable length 1 m
Operating temperature -10 °C to +50 °C
Storage temperature - 30 ºC to +60 ºC
Relative humidity =90 % at 10 °C to 530 °C;
=75 % at 30 °C to 540 °C;
=45 % at 40 °C to 50 °C
IP rating IEC 60529: IP30 when jaw closed
Temperature coefficients Add 0.1 x specified accuracy for each °C above 28°C or below 18 °C.
Operating altitude 2,000 meters
Storage altitude 12,000 meters
Vibration requirements MIL-PRF-28800F, Class 2
Drop test requirements -10 °C, 1m, 6 sides, oak wood
Jaw impact test requirements Follow EN/IEC 61010-2-032:2012
EMI, RFI, EMC IEC/EN 61326-1
Dimensions (hxwxd) 146 mm x 62 mm x 34 mm
Weight 150 g

Please leave your enquiry here, we will reply as soon as possible.


*only support gif, jpeg, jpg, png, pdf
Switch To Desktop Version